medical MODERATE Updated 2026-10-21

Radiometer Medical ApS recalls The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended fo

See all recalls from Radiometer Medical ApS →

Recalled Product

The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order.

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O. ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for The ABL90 FLEX and →
View official government recall