ICU Medical, Inc. recalls 9.5" (24 cm) Appx 0.45 ml, Smallbore Pressure Infusion (400
Recalled Product
9.5" (24 cm) Appx 0.45 ml, Smallbore Pressure Infusion (400 psig) Ext Set w/MicroClave, 2 NanoClave, NanoClave T-Connector, 3 Purple Clamps, Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709049496(17)250101(30)50(10)4603265. The high pressure extension set is a device used as conduit tubing to deliver fluids from an ~f electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices.
Hazard / Issue
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.