medical Class I Updated 2023-11-15

Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P19

Recalled Product

EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190


Hazard / Issue

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model BF-P190 UDI-DI: 4953170342110
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →