medical Class II Updated 2024-10-30

Smiths Medical ASD, Inc. recalls medex TranStar DPT w/Stopcock & 30cc Flush 10/EA, Product Co

Recalled Product

medex TranStar DPT w/Stopcock & 30cc Flush 10/EA, Product Code REF MX950SCNT


Hazard / Issue

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: UDI/DI 50351688503938 (case), 10351688503930 (pouch), Lot Numbers: 4293126, 4309186, 4316649, 4323123, 4328467,4346049