medical Class II Updated 2024-10-30

Smiths Medical ASD, Inc. recalls medex TranStar Stopcock w/Ext. 10/EA, Product Code REF MXG30

Recalled Product

medex TranStar Stopcock w/Ext. 10/EA, Product Code REF MXG300242


Hazard / Issue

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: UDI/DI 50351688517782 (case), 10351688517784 (pouch), Lot Numbers: 4346040, 4385749, 4400278, 4440647