medical Class II Updated 2023-11-08

Nova Biomedical Corporation recalls Nova Biomedical Prime Plus Analyzer -Intended to measure cer

Recalled Product

Nova Biomedical Prime Plus Analyzer -Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 57400 Stat Profile Prime Plus Analyzer (2) PN 59508 Stat Profile Prime Plus Analyzer, Remanufactured


Hazard / Issue

Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: (1) 00385480574006 (2) 00385480595087 All software versions (V1.17.1123.0 and V3.2.1123.0)
View official government recall