medical Class I Updated 2024-11-06

Percussionaire Corporation recalls Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). compo

Recalled Product

Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy


Hazard / Issue

As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.

Issued by

FDA

Affected States: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, UT, ID, ME, MN, SD, ND, KS, KY, TN, MD, VT, MA, RI
Lot/Code Info: Model Numbers: P5-10 P5-HC P5-UC-10 UDI-DI codes: Lot Numbers: 240321 240329 240412 240430 240509 240516 240614 240625 240627
View official government recall