medical Class II Updated 2019-10-30

Beckman Coulter Inc. recalls Beckman Coulter Uric Acid, REF OSR6098, OSR6198, and OSR6698

Recalled Product

Beckman Coulter Uric Acid, REF OSR6098, OSR6198, and OSR6698.


Hazard / Issue

NAPQI, a metabolite of Acetominophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots