medical Class II Updated 2018-10-31

Ecolab Inc recalls Disposable Accessory Kit, 3-Arm (Box of 5) individually seal

Recalled Product

Disposable Accessory Kit, 3-Arm (Box of 5) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the Intuitive Surgical Da Vinci Si Robotic system / Distributed by Intuitive Surgical Product Usage: Microtek Medical Equipment Drapes are intended to cover/wrap various surgical and/or nonsurgical instruments/equipment [e.g. robotic arms, microscopes, tables, x ray systems, light handles, etc]. They function as a physical barrier to prevent cross contamination between the instrument / equipment and medical staff and/or to allow the instrument/equipment to enter a hygienic area [e.g. sterile surgical field] in various clinical settings. They are typically made of flexible plastic and shaped to fit the type of instrument / equipment. They are a single use device [not to be reprocessed]


Hazard / Issue

The pouches of certain lots of product may have wrinkles along the pouch seal that could result in a channel within the seal. Should there be a channel in the pouch seal it may result in a breach in the sterility of the product.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Item 420290-03 Lots # D173047, D173557, D173017, D180127, D172917, D180137, D172827, D172897, D180197A, D172757, D173037, D173557A, D180297, D172907, D180127A, D173207, D172847, D180397, D180037, D180287, D180267, D180197, D180277, D180417, D172787, D180047, D173057, D180407, D172797, D173567, D180477, and D180607.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →