medical Class II Updated 2024-11-06

Abbott Molecular, Inc. recalls Abbott Alinity m STI AMP Kit, used with the Alinity m System

Recalled Product

Abbott Alinity m STI AMP Kit, used with the Alinity m System, product codes: a) REF 09N17-095; b) REF 09N17-090; c) REF 09N17-091;


Hazard / Issue

Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a) REF 09N17-095, UDI/DI 00884999048591; b) REF 09N17-090, UDI/DI 00884999047945; c) REF 09N17-091, UDI/DI 00884999049277 All Lots
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →