medical Class II Updated 2020-10-28

ICU Medical, Inc. recalls 11" (28 cm) Appx 1.7 ml, Ext Set w/6-Port NanoClave Manifold

Recalled Product

11" (28 cm) Appx 1.7 ml, Ext Set w/6-Port NanoClave Manifold, Check Valve, Clamp, NanoClave 4-Way Stopcock, Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709069524(17)250201(30)50(10)4711196. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).


Hazard / Issue

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number:4711196.
View official government recall