medical Class II Updated 2021-12-01

Philips Healthcare recalls Philips Azurion systems with software releases 2.1(L1) and 2

Recalled Product

Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)


Hazard / Issue

3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Azurion 2.1(L1) and 2.1(L2)
View official government recall