medical Class II Updated 2020-10-28

ICU Medical, Inc. recalls 120 IN (305cm)APPX 15.4ml, 10 DROP, 2 Clave STOPCOCK, NanoCl

Recalled Product

120 IN (305cm)APPX 15.4ml, 10 DROP, 2 Clave STOPCOCK, NanoClave T-Connector, 2 CHECK VALVES. 1 unit per pouch, 50 pouches per case. UDI:(01)10887709040141(17)250101(30)25(10)4573652. The SmartValve is a single use, sterile, non-pyrogenic pressure activated check valve system intended for use as an accessory to Intravascular administration set.


Hazard / Issue

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number:4573652.
View official government recall

Looking for safe alternatives?