medical Class II Updated 2020-10-28

ICU Medical, Inc. recalls 22" (56 cm) Appx 2.5 ml, Ext Set w/4-Way Clave Stopcock, Cla

Recalled Product

22" (56 cm) Appx 2.5 ml, Ext Set w/4-Way Clave Stopcock, Clamp, Luer Lock. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709051967(17)250101(30)50(10)4605591; (01)10887709051967(17)250201(30)50(10)4727498. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).


Hazard / Issue

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers:4605591, 4727498.