medical Class II Updated 2024-11-06

Princeton Biomeditech Corp recalls Osom Ultra hCG Combo Test- IVD immunoassay for the qualitati

Recalled Product

Osom Ultra hCG Combo Test- IVD immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine for the early confirmation of pregnancy. Model/Catalog Number: 1004


Hazard / Issue

This recall was initiated due to customer complaints of false negative of hCG test results when running patient urine samples. Use of the above listed product lot may result in false negative results. An end user reported that the OSOM Ultra hCG Combo Test, lot 563L13, produced false negative test results from 2 patient urine samples as confirmed by quantitative hCG and ultrasound. During further root cause testing and investigation, an additional 6 devices were found to produce false negative test results from hCG positive samples.

Issued by

FDA

Affected States: MO
Lot/Code Info: UDI-DI: 00742860100208 Lot Number: 563L13
View official government recall