medical Class II Updated 2020-10-28

ICU Medical, Inc. recalls 8" (20 cm) Appx 0.33 ml, Smallbore Ext Set w/MicroClave Clea

Recalled Product

8" (20 cm) Appx 0.33 ml, Smallbore Ext Set w/MicroClave Clear, NanoClave Clamp, Rotating Luer.1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709062946(17)250101(30)50(10)4603254; (01)10887709062946(17)250201(30)50(10)4630684; (01)10887709062946(17)250201(30)50(10)4718516; (01)10887709062946(17)241201(30)50(10)4514832; (01)10887709062946(17)241201(30)50(10)4537159; (01)10887709062946(17)250101(30)50(10)4558447; (01)10887709062946(17)250201(30)50(10)4724820; (01)10887709062946(17)250201(30)50(10)4733554. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).


Hazard / Issue

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers:4603254, 4630684, 4718516, 4514832, 4537159, 4558447, 4724820, 4733554.
View official government recall