medical Class II Updated 2021-12-01

Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Op

Recalled Product

Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 64 System Number: 882471 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.


Hazard / Issue

Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Number: 7141 7218 7229 7245
View official government recall