medical Class II Updated 2018-11-07

Magellan Diagnostics, Inc. recalls Magellan LeadCare¿ II Blood Lead Test Kit. In Vitro Test for

Recalled Product

Magellan LeadCare¿ II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood. Catalog Number: 70-6762 Primary DI No: 00850355006000 The LeadCare II Blood Lead Test Kit is for in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and treatment reagent of the test kit are specific for lead only. Contents of the LeadCare II Blood Lead Test Kit must only be used with a LeadCare II Analyzer.


Hazard / Issue

Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 ¿g/dL). Results for patient testing with this lot will be displayed as HIGH.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot 1808M and sub-lots: 1808M-01, 1808M-02, 1808M-03, 1808M-04, 1808M-05, 1808M-06, 1808M-07, 1808M-08, 1808M-09. Exp. Date:November 10, 2019
View official government recall