medical Class II Updated 2023-11-15

Maquet Medical Systems USA recalls Cardiohelp-i, blood oxygenation and carbon dioxide removal s

Recalled Product

Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)


Hazard / Issue

It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →