medical Class II Updated 2018-11-07

Siemens Medical Solutions USA, Inc recalls Multix Fusion system is a radiographic system: Models: Mul

Recalled Product

Multix Fusion system is a radiographic system: Models: Multix Fusion Analog (model number 10746665) with Tube stand 3D V (model number 7042042 or 7042059); Multix Fusion Digital Portable (model number 10746666) with Tube stand 3D V (model number 7042042); Multix Fusion Digital Integrated (model number 10746667) with Tube stand 3D V (model number 7042042); Multix Fusion Digital Wireless (model number 10893300) with Tube stand 3D V (model number 7042042)


Hazard / Issue

Steel cables inside the lifting column of the overhead tube, may rapture without triggering the safety lock, leading to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 40279 10037 1176 40026 1220 40106 40208 40209 40109 40114 40140 40104 40126 40144 10040 10041 40027 40118 40105 40128 40200 40146 10039 40102 40101 40022 40125 40142 40127 40117 40206 1192 1185 40116 40119