medical Class II Updated 2018-11-07

Siemens Medical Solutions USA, Inc recalls AXIOM Multix MT (model number 8395415 or 8395399 or 8395381)

Recalled Product

AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Systems radiographic exposures of the entire body.


Hazard / Issue

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.

Issued by

FDA

Affected States: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WI, WY
Lot/Code Info: Serial Numbers 1137 1138 1525 1121 1132 1529 1074 1531 1472 1136 1119
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →