medical Class II Updated 2021-12-08

BioFire Diagnostics, LLC recalls BioFire Blood Culture Identification 2 (BCID2) Panel Part No

Recalled Product

BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All lots with an expiry date prior to


Hazard / Issue

Due to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results. The cause may be as a due to non-viable contamination in the blood culture vials.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Blood Culture Identification 2 (BCID2) Panel Part Number: RFIT-ASY-0147 UDI Code: 00815381020338 Blood Culture Media Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All BCID2 Panel lots BD Blood Culture Media vials with expiration date of April 30, 2022 or earlier.