medical Class II Updated 2023-11-29

Medtronic Inc recalls Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, M

Recalled Product

Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5


Hazard / Issue

The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI/DI 08033477090696:
View official government recall