medical Class II Updated 2023-11-29

Sentinel CH SpA recalls IRON assay, Reference Numbers 6K95-41 and 6K95-30

Recalled Product

IRON assay, Reference Numbers 6K95-41 and 6K95-30


Hazard / Issue

A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Reference Number 6K95-41: UDI-DI: 00380740010928; Lot Number: 30610Y600; Reference Number 6K95-30: UDI-DI: 00380740010911; Lot Numbers: 21058Y600, 21101Y600, 30918Y600