Sentinel CH SpA recalls IRON assay, Reference Numbers 6K95-41 and 6K95-30
Recalled Product
IRON assay, Reference Numbers 6K95-41 and 6K95-30
Hazard / Issue
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
Issued by
FDA