medical Class II Updated 2023-11-29

Sentinel CH SpA recalls Alinity c Iron Reagent, Reference Number 08P3920

Recalled Product

Alinity c Iron Reagent, Reference Number 08P3920


Hazard / Issue

A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00380740136253; Lot Number: 21046Y600, 21118Y600, 30227Y600, 30302Y600
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →