medical MODERATE Updated 2026-11-11

Baxter Healthcare Corporation recalls TruSystem 7000 U Operating Table (dV), Product Code 1723633

See all recalls from Baxter Healthcare Corporation →

Recalled Product

TruSystem 7000 U Operating Table (dV), Product Code 1723633

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for TruSystem 7000 U Operating →
View official government recall