medical Class II Updated 2023-11-29

Aizu Olympus Co., Ltd. recalls Hysteroscope-For endoscopic diagnosis within the uterus Mod

Recalled Product

Hysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.


Hazard / Issue

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number (UDI-DI): HYF-1T (04953170051128).
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →