medical Class II Updated 2023-11-29

Aizu Olympus Co., Ltd. recalls Hysterovideoscope-For endoscopic diagnosis within the uterus

Recalled Product

Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.


Hazard / Issue

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number (UDI-DI): HYF-V (04953170340185).
View official government recall

Looking for safe alternatives?