medical Class II Updated 2023-11-29

Aizu Olympus Co., Ltd. recalls Tracheal fiberscope-airway management, which includes observ

Recalled Product

Tracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: LF-DP, LF-GP, LF-TP.


Hazard / Issue

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number (UDI-DI): LF-DP (04953170136856), LF-GP (04953170061998), LF-TP (04953170317576).