medical Class II Updated 2023-11-29

Aizu Olympus Co., Ltd. recalls Mobile airway scope-diagnosis and observation to access airw

Recalled Product

Mobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-TM2.


Hazard / Issue

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Numbers: MAF-DM2 (04953170407260), MAF-GM (04953170340369), MAF-GM2 (04953170407208), MAF-TM (04953170288630), MAF-TM2 (04953170407147).