medical Class II Updated 2023-11-29

Aizu Olympus Co., Ltd. recalls Ultrasonic gastrovideoscope-Real time ultrasound imaging,ult

Recalled Product

Ultrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs Model Number: TGF-UC180J.


Hazard / Issue

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number (UDI-DI): TGF-UC180J (04953170328480).
View official government recall

Looking for safe alternatives?