medical Class II Updated 2020-11-11

Zimmer Biomet, Inc. recalls ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethyle

Recalled Product

ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 57 mm Number: 00-8065-554-26


Hazard / Issue

Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots
View official government recall