medical Class II Updated 2023-11-29

Aizu Olympus Co., Ltd. recalls Ureteralscope- for endoscopic diagnosis and treatment within

Recalled Product

Ureteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). Model Numbers: URF-P6, URF-P6R.


Hazard / Issue

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number (UDI-DI): URF-P6 (04953170340802), URF-P6R (04953170340833).