Aizu Olympus Co., Ltd. recalls Uretero-reno fiberscope-For endoscopic diagnosis and treatme

Recalled Product

Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.


Hazard / Issue

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number (UDI-DI): URF-P7 (04953170403811), URF-P7R (04953170403835), URF-V2 (04953170343582), URF-V2R (04953170343612), URF-V3 (04953170403385), URF-V3R (04953170403392).