medical MODERATE Updated 2026-11-11

Baxter Healthcare Corporation recalls Operating Table TruSystem 7000 U (dV), Model/Catalog Number:

See all recalls from Baxter Healthcare Corporation →

Recalled Product

Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial numbers: 108920104, 110192607, 110192885, 110208936, 110212885, 110220066, 110227520, 110248337, 110250614, 109423464, 110192608, 110192886, 110208941, 110212886, 110220068, 110227523, 110248343, 110250621, 109517138, 110192609, 110206656, 110210717, 110220055, 110220069, 110227524, 110249706, 110250622, 110139767, 110192683, 110206657, 110210718, 110220056, 110220071, 110227525, 110249712, 110206662, 110139776, 110192882, 110206659, 110210720, 110220057, 110220072, 110227552, 110250605, 110188921, 110188919, 110192883, 110206660, 110210721, 110220058, 110220078, 110227553, 110250609, 110188922, 110188920, 110192884, 110206664, 110212881, 110220065, 110220079, 110248218, 110250613; UDI-DI number: 00887761967519

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
View official government recall