medical Class II Updated 2019-11-20

Canon Medical System, USA, INC. recalls Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table

Recalled Product

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.


Hazard / Issue

Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model CAT-870B D053991 Serial No. Manufacture Date 99D08X2101 October 2008 99D0582039 August 2005 99D0762072 June 2007 99D0852090 May 2008 99D07Y2078 November 2007 99D0812081 January 2008 99D0582042 August 2005 99D0552033 May 2005 D551206 May 2005