medical Class II Updated 2021-12-22

Smith & Nephew, Inc. recalls smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, RE

Recalled Product

smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, REF 72573101


Hazard / Issue

Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Number: 20HM10249, 20HM10250, 20HM10251, 20HM22783, UDI (01) 00885556752999 (17) 310630 (10) 12345678 (11) 2210630 (21) 87654321
View official government recall