medical Class II Updated 2018-11-28

GE Healthcare Biosciences recalls DeltaVision OMX SR Imaging System Model Number: 29115476

Recalled Product

DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laboratory equipment for super resolution imaging offluorescently-la beling biological samples. Techniques such as 3D Structured Illumination, Total Internal Reflection Fluorescence and localization microscopy may be used to examine a variety of samples over long periods of time.


Hazard / Issue

Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: System Serial number OM20072 - OM40122
View official government recall