medical Class II Updated 2023-11-29

LEONI CIA CABLE SYSTEMS recalls ORION System with software OSS v6.0 and OSS v7.0- An electro

Recalled Product

ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00003


Hazard / Issue

Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: (01)03700383510017 SN: 100000001141
View official government recall