medical Class II Updated 2021-12-29

Maquet Cardiovascular, LLC recalls Hemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference

Recalled Product

Hemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890


Hazard / Issue

There is a probable mislabeling of product. A Hemashield Platinum Woven Double Velour Bifurcated Vasuclar Graft, SN 1259735369, 24 mm x 12 mm, REF Number M00202166241P0 is likely packaged inside of the carton instead of a Hemashield Gold Knitted Microvel Double Velour Bifurcated Vascular Graft; SN 1260004700; 18 mm x 9 mm; REF Number M002020851890 or a Hemashield Gold Knitted Microvel Double Velour Bifurcated Vascular Graft; SN 1259882880; 18 mm x 9 mm; Ref Number M002020851890.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot number: 20C04; Serial Numbers: 1260004700 and 1259882880; UDI Code: (01)00384401017752(11)200304(17)250228(21)1260004700GS(10)20C04 (01)00384401017752(11)200304(17)250228(21)1259882880GS(10)20C04
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →