medical Class II Updated 2020-11-11

Cardinal Health 200, LLC recalls Covidien MONOJECT Standard Hypodermic Needle, Polypropylene

Recalled Product

Covidien MONOJECT Standard Hypodermic Needle, Polypropylene Hub 22 G x 3/4". Intended use to facilitate activities such as: withdrawal of medication from a container, aspiration of fluid from a patient, injection of medication into a patient, and transfer fluids Item Code: 8881250248


Hazard / Issue

Potential for the needle to detach from the cartridge after the bottom of the cartridge is twisted and removed. If the needle is not securely seated, the needle can fall out of the sheath and cause the unused needle to project out of the cartridge, and result in a needle stick/puncture from the clean/unused needle

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number: 011144 Expanded Recall 11/4/20: Lot Number: 011145
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →