medical Class II Updated 2020-11-18

Fresenius Medical Care Holdings, Inc. recalls Liberty Select Cycler Software Version 2.9.0 Model Number

Recalled Product

Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343


Hazard / Issue

The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra fluid is present and it will not make it available for treatment. Not having enough solution to complete treatment may lead to inadequate dialysis which could result in the accumulation of toxins and fluid. Symptoms may include shortness of breath, increased weight gain, and increased blood pressure.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All Liberty Select Cyclers with Software Version 2.9.0 Model Number 180343 and RTLR180343
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →