Ge Inspection Technologies Llc recalls v|tome|x L Series non-destructive testing cabinet x-ray syst
Recalled Product
v|tome|x L Series non-destructive testing cabinet x-ray systems
Hazard / Issue
It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.