medical Class II Updated 2022-01-05

AMO Manufacturing USA, LLC recalls Catalys Precision Laser System, Models: Catalys-U (United St

Recalled Product

Catalys Precision Laser System, Models: Catalys-U (United States), Catalys-I (International), Catalys-C (China); and Catalys Precision Laser System No Bed, Models: 0160-6020 (United States), 0160-6010 (International)


Hazard / Issue

Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.

Issued by

FDA

Affected States: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY, IN, CO, NJ, OK, NH, UT, AR, NY, AK, ND, NE, IL, AL, HI, CT, DE, KS, GA, OR, MS, PA, MT, ME, MI, AZ, ID, WI, VT, NM
Lot/Code Info: Catalys-U and Catalys-I with Windows 7 Ultimate/Professional/Embedded Standard. Catalys-C with Windows 7 Embedded Standard. 0160-6020 and 0160-6010 with Windows 7 Professional/Embedded Standard.
View official government recall