medical Class II Updated 2024-11-20

AVID Medical, Inc. recalls Halyard IVR VASCULAR PACK - Medical convenience kits Model

Recalled Product

Halyard IVR VASCULAR PACK - Medical convenience kits Model Number: VAST027-08


Hazard / Issue

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN: 10809160416303 Lot Number: 1608916, 1608917
View official government recall