medical Class II Updated 2022-01-12

Invacare Corporation recalls Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiN

Recalled Product

Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: IFX-20R, IFX-20SP, IFX-20MP, IFX-20C


Hazard / Issue

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: IFX-20C 00841447111627 IFX-20MP 00841447111658 IFX-20R 00841447111634 IFX-20SP 00841447111641 IFX20WIDE 00841447111658