medical Class II Updated 2018-11-21

Siemens Medical Solutions USA, Inc recalls ARTIS zee biplane systems with Material 10094141- a Angiogra

Recalled Product

ARTIS zee biplane systems with Material 10094141- a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee includes cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee can also support the acquisition of position triggered imaging for spatial data synthesis.


Hazard / Issue

Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: Serial 155122 155149 155151 155105 155148 154671 155126 155131 155132 155119 155121 155118 155147