medical Class II Updated 2024-11-27

Bard Peripheral Vascular Inc recalls Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MA

Recalled Product

Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF: MQ1210; BARD MARQUEE Kit 12G X 13 CM, REF: MQK1213; BARD MARQUEE 12G X 13 CM, REF: MQ1213


Hazard / Issue

Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacement, which could lead to tissue damage and bleeding.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: REF/UDI-DI/Lot(Expiration): MQK1210/00801741097263/1490758(9/28/2025), 1506044(1/28/2026), 1506045(1/28/2026), 1511827(2/28/2026), 1514692(3/28/2026), 1518977(4/28/2026), 1518978(4/28/2026), 1518979(4/28/2026), 1519143(4/28/2026), 1520507(5/28/2026), 1520508(5/28/2026), 1522157(5/28/2026), 1522158(5/28/2026), 1522577(5/28/2026), 1523785(6/28/2026), 1523914(6/28/2026), 1523915(6/28/2026), 1525370(6/28/2026), 1525371(6/28/2026), 1525373(6/28/2026), 1528417(7/28/2026), 1528418(7/28/2026), 1528495(7/28/2026), 1528496(7/28/2026), 1528497(7/28/2026), 1528560(7/28/2026), 0001618867(3/28/2028), 0001620281(3/28/2028), 0001632675(6/28/2028); MQ1210/00801741097119/1514460(3/28/2026), 1514461(3/28/2026), 1519144(4/28/2026), 1519145(4/28/2026), 1519456(4/28/2026), 1520066(5/28/2026), 1520067(5/28/2026), 1520506(5/28/2026), 1523783(6/28/2026), 1523784(6/28/2026), 0001618258(3/28/2028), 0001631611(6/28/2028), 0001633216(7/28/2028); MQK1213/00801741097270/1543517(10/28/2026), 1559312(2/28/2027), 1562724(3/28/2027), 1569737(4/28/2027), 1581652(7/28/2027); MQ1213/00801741097126/1558191(2/28/2027), 1559313(2/28/2027)
View official government recall

Looking for safe alternatives?