medical Class II Updated 2020-11-25

Siemens Medical Solutions USA, Inc recalls Artis zee floor, Artis Q ceiling, Artis Q.zen ceiling. Image

Recalled Product

Artis zee floor, Artis Q ceiling, Artis Q.zen ceiling. Image-Intensified Fluoroscopic X-Ray System. Material numbers 10848281, 10848354, 10094135, 10848354 - Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.


Hazard / Issue

Some electronic components within the generator were equipped with an incorrect protection diode. As a result, protection against overvoltage may be inadequate, and the system may not function properly in the event an overvoltage situation occurs.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial numbers 109762 111158 124344 111161
View official government recall