medical Class II Updated 2024-11-27

Hobbs Medical, Inc. recalls Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723

Recalled Product

Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723


Hazard / Issue

Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog Number: 4723; UDI-DI: M84947230; Lot Numbers: H07-23-107, H08-23-034.